CM101-Dualer-IQ-Pro

 Dualer IQ Pro™ Digital Inclinometer

The Dualer IQ Pro Digital Dual Inclinometer is the faster, easier, smarter way to measure spine and extremity ranges of motion without a computer. Dualer IQ Pro speeds up spine evaluations with dual inclinometry protocols recommended by the American Medical Association (AMA) and automatic data collection. Dualer IQ Pro produces easy to read digital results. The Dualer IQ Pro records and stores results for up to 20 tests allowing clinicians to perform consecutive tests without stopping to record data. Later review of test data allows for less distractions and less transcription errors. Dualer IQ Pro allows clinicians to evaluate motion using dynamic dual Inclinometry and static single Inclinometry (similar to using Goniometer) protocols.

Generally, this is unnecessary as the Dualer IQ Pro comes fully calibrated from JTECH.

Follow These Steps to Perform a Full Calibration:

Note: The angles for a full calibration on the inclinometers must be exact for the calibration to be valid.

It is recommended to use a T-Square or other tools that have exact right angles. You should also always
use the same surface on the inclinometer for reference; the feet and the back of the inclinometer are the
recommended reference surfaces on the inclinometer
1. Begin with the Secondary inclinometer disconnected from the Primary inclinometer
2. Select “Settings” from the main menu.
3. Use the Up or Down buttons until you highlight “Calibration.”
4. Press the Select Button to begin calibration.
5. Set the inclinometer in the upright position.
6. Press the Select Button to save this position.
7. Rotate the inclinometer 90 degrees to the left.
8. Press the Select Button to save this position.
9. Rotate the inclinometer 90 degrees more to the left, so it is 180 degrees from the starting position.
10. Press the Select Button to save this position.
11. Rotate the inclinometer an additional 90 degrees to the left, so it is 270 degrees from the starting position.
12. Press the Select Button to save this position.
13. Place the back of inclinometer on a flat surface with the LCD face up.
14. Press the Select Button to save this position.
15. Place the front of the inclinometer on a flat surface with the LCD facing down.
16. Press the Enter Button to save this position.
17. Connect the Secondary inclinometer and repeat the process
Note: For information on recommended annual factory recalibration, contact JTECH Customer Service at (800) 985-8324, ext. 4 or This email address is being protected from spambots. You need JavaScript enabled to view it.